Treatment of ruptured intracranial aneurysms with the Woven EndoBridge device: one-year angiographic and clinical follow-up in a single-center cohort
Article information
Abstract
Objective
The Woven EndoBridge (WEB) device is an intrasaccular flow disruptor for wide-neck bifurcation aneurysms, avoiding adjunctive stents or dual antiplatelet therapy. Evidence in ruptured cases remains limited. We evaluated 1-year angiographic and clinical outcomes of ruptured aneurysms treated with the WEB device.
Methods
Thirteen patients with aneurysmal subarachnoid hemorrhage treated between January 2023 and December 2024 were retrospectively analyzed. Hunt-Hess grade, modified Fisher grade, and 12-month modified Rankin Scale (mRS) scores were reviewed. Follow-up included 6-month magnetic resonance angiography and 12-month digital subtraction angiography (DSA). Occlusion was classified via the WEB Occlusion Scale (WOS), with adequate occlusion defined as WOS A or B.
Results
The cohort comprised 11 males and two females (mean age, 49.4 years; range, 41–82). Hunt-Hess grades were II in 69% (9/13), IV in 23% (3/13), and V in 8% (1/13). Modified Fisher grades were 3 in 38% (5/13) and 4 in 62% (8/13). At 12-month DSA, adequate occlusion was achieved in 69% (9/13; WOS A: n=3, WOS B: n=6). Neck remnants occurred in three aneurysms (23%), and one aneurysm remnant (8%) required stent-assisted coiling. Twelve-month mRS scores were 0 in nine patients (69%), 3 in two (15%), and 4 in two (15%). No complications, rebleeding, or deaths occurred.
Conclusion
Primary WEB embolization of ruptured aneurysms achieved favorable 1-year occlusion without rebleeding, demonstrating feasibility when avoiding stents and antiplatelet therapy is desirable.
INTRODUCTION
Endovascular treatment is recognized as the first therapeutic option for ruptured and unruptured intracranial aneurysms [1]. Endovascular treatment of wide-neck bifurcation aneurysms is technically challenging, and these aneurysms are associated with a higher rate of recanalization [2]. Bridging-neck devices such as stents require dual-antiplatelet therapy and are associated with a higher rate of complications; in ruptured aneurysms, this anti-aggregation regimen carries an inherent risk of early rebleeding and hemorrhagic complications in the perioperative period [3-5]. Despite anti-aggregation, thromboembolic complications still occur [6,7].
The Woven EndoBridge (WEB; MicroVention) is an intrasaccular flow disruptor developed primarily for the endovascular treatment of wide-neck bifurcation aneurysms, without the need for adjunctive stents or dual antiplatelet therapy. The first clinical results of the WEB device demonstrated good safety and efficacy [8]. However, published data on WEB treatment of ruptured aneurysms remain limited, and follow-up in most series does not extend beyond 1 year on average [9,10]. A single-center series of 100 ruptured aneurysms treated primarily with the WEB has been reported, with favorable long-term results [11,12].
In this study, we report the 1-year angiographic and clinical follow-up of a consecutive single-center cohort of ruptured intracranial aneurysms treated with the WEB device as primary treatment.
MATERIALS AND METHODS
Study design and population
This retrospective observational study analyzed a prospectively maintained institutional database and was conducted in accordance with our institutional privacy policy; institutional review board approval (Hallym University Dongtan Sacred Heart Hospital, No. HDT-2023-11-008) and the requirement for informed consent were handled per institutional policy. From January 2023 through December 2024, 122 cases of ruptured intracranial aneurysms were treated at our center: 14 by surgical clipping, 76 by conventional coil embolization, 19 by stent-assisted coil embolization, and 13 by the WEB device. All 13 consecutive WEB-treated patients constituted the final analytic cohort (Fig. 1).
Woven EndoBridge device
The WEB is a self-expanding, retrievable, electrothermally detachable, nitinol braided device placed within the aneurysm sac. Current single-layer iterations are available as a barrel-shaped WEB Single-Layer (SL) and a spherical WEB Single-Layer Sphere, with enhanced-visualization variants incorporating composite nitinol-platinum wire strands. Once deployed, the WEB modifies blood flow at the aneurysm neck and induces intra-aneurysmal thrombosis. The device can be fully retrieved until final electrothermal detachment. In this cohort, all aneurysms were treated with a WEB SL device; device sizes ranged from 4.0×2 mm to 10×6 mm (Table 1). Device width was selected 1–2 mm larger than the maximal aneurysm width, and device height was selected approximately 1 mm shorter than the measured aneurysm height.
Indications and treatment
Treatment modality and technique were determined by multidisciplinary discussion, and ruptured aneurysms were treated acutely after admission. The WEB was favored for aneurysms suitable for intrasaccular flow disruption when avoidance of an adjunctive stent and dual antiplatelet therapy was desirable. With the patient under general anesthesia and via a transfemoral approach, a microcatheter was navigated into the aneurysm through a coaxial system with an intermediate catheter. An intravenous bolus of 3,000 IU unfractionated heparin was administered at the start of the procedure. Because all procedures lasted less than 60 minutes, no additional heparin was administered; activated clotting time was not routinely measured. After WEB implantation, aspirin 100 mg once daily was prescribed for 2 weeks; dual antiplatelet therapy was not used. In one patient (No. 12), adjunctive coiling was performed in addition to the WEB.
Data collection and follow-up
Patient demographics, baseline Hunt-Hess and modified Fisher grades, aneurysm characteristics (maximal dome width, height/depth, and neck width), treatment details, procedure-related complications, rebleeding, retreatment, mortality, and 12-month modified Rankin Scale (mRS) scores were collected. All patients underwent magnetic resonance angiography (MRA) at 6 months for interval surveillance and catheter digital subtraction angiography (DSA) at 12 months for the primary angiographic endpoint. Twelve-month DSA was interpreted by two experienced neurointerventionists. Occlusion was graded using the WEB Occlusion Scale (WOS): complete occlusion (WOS A), complete occlusion with residual opacification of the proximal recess of the device (complete occlusion with proximal-recess residual; WOS B), neck remnant (WOS C), and aneurysm remnant (WOS D). Adequate occlusion was defined as WOS A or B. A favorable clinical outcome was defined as an mRS score of 0–2.
Statistical analysis
Quantitative variables were expressed as means and medians with ranges; qualitative variables were expressed as frequencies and percentages. Given the small sample size, analyses were descriptive.
RESULTS
Population
From January 2023 through December 2024, 13 of 122 ruptured intracranial aneurysm cases were treated with the WEB device and included in this analysis. There were 11 males and two females, with a mean age of 49.4 years (median, 44 years; range, 41–82 years). All 13 patients presented with subarachnoid hemorrhage. Hunt-Hess grades were II in nine patients (69%), IV in three (23%), and V in one (8%); modified Fisher grades were 3 in five patients (38%) and 4 in eight (62%) (Table 2). Aneurysm locations were the anterior communicating artery in six patients (46%), the middle cerebral artery bifurcation in three (23%), and the basilar tip in four (31%).
The mean maximal aneurysm diameter was 6.3 mm (median, 5.5 mm; range, 3.6–10.3 mm), and the mean neck width was 3.9 mm (median, 3.4 mm; range, 2.6–7.3 mm). Seven aneurysms (54%) met the definition of a wide neck. All aneurysms were treated with a WEB SL device, with sizes ranging from 4.0×2 mm to 10×6 mm. One patient (No. 12) underwent adjunctive coiling in addition to the WEB (Table 1).
Follow-up
All 13 patients completed the scheduled 6-month MRA and 12-month DSA examinations. Twelve-month clinical follow-up and mRS scores were also available for all patients.
Morbidity and mortality
There were no procedure-related complications, post-treatment thromboembolic events, rebleeding, or deaths during follow-up. At 12 months, mRS scores were 0 in nine patients (69%), 3 in two (15%), and 4 in two (15%). A favorable outcome (mRS 0–2) was observed in nine of 13 patients (69%) (Table 2).
Aneurysm occlusion and retreatment
At 12-month DSA (Table 1), nine of 13 aneurysms (69%) demonstrated adequate occlusion: three (23%) were completely occluded (WOS A) and six (46%) showed complete occlusion with residual opacification limited to the proximal recess of the device (WOS B). Representative cases are shown in Fig. 2 and 3. A neck remnant (WOS C) was present in three aneurysms (23%), and an aneurysm remnant (WOS D) was present in one aneurysm (8%). The single aneurysm remnant occurred in a wide-neck anterior communicating artery aneurysm (No. 7; dome, 8.5 mm; neck, 5.0 mm) and was identified on the 12-month DSA. Stent-assisted coil embolization was performed 2 weeks later (Fig. 4).
(A) Pretreatment digital subtraction angiography (DSA) demonstrating the aneurysm. (B) Post-deployment image showing the Woven EndoBridge (WEB) device in place. (C) Twelve-month follow-up DSA demonstrating complete occlusion of the aneurysm (WEB Occlusion Scale A) with preservation of the parent and branch vessels.
(A) Pretreatment digital subtraction angiography (DSA) in the working projection demonstrating the aneurysm. (B) Immediate post-deployment image showing the Woven EndoBridge (WEB) device positioned within the aneurysm sac. (C) Twelve-month follow-up DSA showing complete occlusion of the aneurysm with residual opacification confined to the proximal recess of the device (WEB Occlusion Scale B) and a patent parent artery.
(A) Pretreatment digital subtraction angiography (DSA) showing the wide-neck aneurysm. (B) Post-deployment image with the Woven EndoBridge (WEB) device within the sac. (C) Follow-up DSA demonstrating an aneurysm remnant with contrast filling of the sac (WEB Occlusion Scale D). (D) Retreatment with stent-assisted coiling, with subsequent angiographic occlusion of the aneurysm and no procedure-related complication.
DISCUSSION
This single-center study demonstrates that primary treatment of ruptured intracranial aneurysms with the WEB device is feasible, with adequate angiographic occlusion (WOS A+B) achieved in 69% of aneurysms at 12-month DSA and no rebleeding during follow-up. A neck remnant was observed in 23% and an aneurysm remnant in 8%. At 12 months, 69% of patients had a favorable clinical outcome (mRS 0–2), and no patient died. More than half of the treated aneurysms (54%) had a wide neck, reflecting the primary indication for the device.
Prevention of rebleeding is the primary goal in the treatment of ruptured aneurysms. In our cohort, there were no rebleeds during follow-up, consistent with prior WEB series in the ruptured setting [9,10] and comparable to reported rates after coiling [13,14]. Patients received aspirin monotherapy at 100 mg daily for only 2 weeks after WEB implantation, and no dual antiplatelet therapy was used. This short antiplatelet regimen is an important advantage compared with stent-assisted coiling in the acute phase of subarachnoid hemorrhage, in which dual antiplatelet therapy may increase hemorrhagic risk [3-5,7]. Two meta-analyses of stent-assisted coiling for ruptured wide-neck aneurysms reported total complication rates of approximately 20% [6,15].
Residual opacification limited to the proximal recess of the device (WOS B) was the most common finding in our cohort. This appearance is frequently observed after WEB implantation and is related to the well-described phenomenon of “WEB shape modification,” a decrease in device height due to deepening of the central recess, probably from compression [16,17]. Such residuals and small neck remnants are generally expected to remain stable and often do not require retreatment; nevertheless, the clinical significance of neck remnants remains debated, because remnants after aneurysm treatment carry an inherent risk of re-rupture, as demonstrated for coiling [18]. In our series, the four aneurysms with a neck or aneurysm remnant tended to have larger dimensions and wider necks, in keeping with the recognized association between aneurysm and neck size and incomplete occlusion; continued angiographic surveillance of these aneurysms is therefore warranted.
Our findings are consistent with the larger single-center experience of van Rooij et al. [11] and Kortman et al. [12], in which primary WEB treatment of 100 ruptured aneurysms produced stable occlusion with a low retreatment rate and no rebleeds on long-term follow-up. The present cohort adds 1-year data from an independent single-center series and supports the WEB as a valuable alternative to coils and to stent-assisted coiling for ruptured wide- and small-neck aneurysms.
Limitations
This study is limited by its single-center, uncontrolled, retrospective design and small sample size (n=13), which preclude robust statistical inference and increase the possibility of selection bias. Angiographic results were assessed at the treating center without independent core-laboratory adjudication. Although all patients underwent 6-month MRA and 12-month DSA, follow-up beyond 1 year was not available; longer-term surveillance is required to confirm the durability of occlusion.
CONCLUSION
In this single-center cohort, primary WEB treatment of ruptured intracranial aneurysms was feasible and achieved adequate occlusion in the majority of aneurysms at 1 year without rebleeding. The WEB may offer an alternative to conventional and stent-assisted coiling without the need for an adjunctive stent or dual antiplatelet therapy; only short-term aspirin monotherapy was used in this cohort. Larger series with longer follow-up are needed to confirm these findings.
Notes
CONFLICTS OF INTEREST
No potential conflict of interest relevant to this article was reported.